PAMAS S40 AVTUR
A wear resistant ceramic piston pump guarantees a constant flow. With a simple-to-use touch screen user interface, operation is easy and intuitive. The system is equipped with an integrated battery for more than three hours operation and a data storage of more than 4000 measurements. The samples can be taken from bottles without pressure or pressurised up to 6 bar.
Portable particle counting system for aviation turbine fuel
According to IP 577 and DEF STAN 91-91
Applications:
•Online measurements at pressure-less systems
•Offline measurement using sample bottle (laboratory mode)
•Long-term analysis
•Bypass filtration monitoring
•Filter verification
Particle sensor: PAMAS HCB-LD-50/50
Max. particle concentration: 24.000 particles per ml at 7.8 % coincidence
Nominal flow rate: 25 ml/min
Calibration according to ISO 4406 in the size classes > 4 µm(c), > 6 µm(c) and > 14 µm(c)
Software:
Download S40: used to transfer stored measurements to a PC,
PAMAS CMDM: Condition Monitoring Data Manager
PAMAS PMA: Particle Measuring and Analysing Software