Applications
- water based hydraulic fluids
- Subsea Christmas tree and wellhead assemblies
- Hydraulic-Power-Units
- Hydraulic Accumulators
- Subsea Umbilicals
- Hydraulic valves and control systems
Standards and methods
ISO 11171 | ISO 4402 | ISO 21501-3 | ISO 4406 | SAE AS4059 | NAS 1638 | ISO 11500 |
Viscosity
max. 1600 cSt with optional high-pressure pump or with external pressure supply
Volumetric particle sensors
PAMAS HCB-LD-50/50
Size range:
1 – 400 µm with calibration according to ISO 21501
1 – 100 µm with calibration according to ISO 4402
4 – 70 µm(c) with calibration according to ISO 11171
Max. particle concentration: 24,000 p/ml at a flow rate of 25 ml/min and a coincidence quote of 7.8%
PAMAS HCB-LD-25/25
Size range:
1 – 200 µm with calibration according to ISO 21501
1 – 100 µm with calibration according to ISO 4402
4 – 70 µm(c) with calibration according to ISO 11171
Max. particle concentration: 120,000 p/ml at a flow rate of 25 ml/min and a coincidence quote of 7.8%