| Specifications
Performance |
|
| Sensitivity | 50 ppt in 3 sec./10 ppt in 60 sec. |
| Noise at zero (1σ) | <25 ppt in 3 sec. |
| Detection limit | 3 ppt |
| Integration interval | selectable: 0.1…999 |
| Rise time (0 – 95%) | <1 sec |
| Zero drift | non (pre chamber) |
| Linearity deviation | <1% full-scale |
| Interferences | HC’s, NH3, NOY non |
| Operating Specifications | |
| Ranges | 5, 10, 50, 100, 500 ppb |
| Outputs | serial: RS232
analog :1V, 10V, at >500 kΩ / 4-20 mA at <600 Ω |
| Temperature range | 5-50°C |
| Humidity tolerance | 5 – 95% rel. humidity |
| Gas flow | sample: 3 l/min NPT
O2: 330 ml/min NPT dry air : <50 ml/min NPT |
| Reaction chamber pressure | 14 mbar |
| PMT cooling temp. | <-15°C |
| Sample inlet temp. | 60°C regulated |
| Operating voltage | standard: 28 VDC ±6%
optional: 24 VDC ±10% |
| Power requirements | 200 W max. |
| Delivery includes | |
| NO/NOX analyzer with all electrical cables, two silica-gel cartridges. | |
| Delivery excludes | |
| Vacuum pump, vacuum tubing and ozone destroyer | |
| Physical characteristics | |
| Dimensions (mm) | casing: width: 440/height: 225/depth: 420
front: width: 483/height: 264/depth: 4 |
| Weight (kg) | 35 |
| Material | standard: aluminum
optional: aerospace aluminum |
| Connections | all connections situated on front panel
28 VDC 1x RS232 3x analog output Connection for PLC O2 inlet (1/4” Swagelok) dry air inlet (1/4” Swagelok) sample inlet (1/4” Swagelok) vacuum outlet (DN 16 ISO KF) |
| Options | |
| Pressure regulation | inlet pres. reg. system (bypass concept) |
| Increased sensitivity | 25 ppt in 3 sec. /· 5 ppt in 60 sec |
| NO2 converter | PLC 762 SR (photolytic converter) |
| ECO PHYSICS reserves the right to change these specifications without notice. | |